Performance Characteristics Physical Characteristics


Nominal Range Max OverloadExpected Operating Life


Output Signal

0-25% Oxygen

30% Oxygen Two years in air

0.195 - 0.25mA in air

Storage Life Recommended Storage

Temperature

Warranty Period

Six months in CTL container 0-20°C

 

24 months from date of despatch

(This amounts to a variation of condition


T95

Response Time

£15 seconds

6 of our standard terms and conditions which otherwise apply)


Offset (3mins N2) Temperature Range Temperature Coefficient Absolute Pressure Range

Differential Pressure

Range Pressure Coefficient

Operating Humidity

intermittent

<0.5% O2

-20°C to +50°C

0.2% signal/°C

Atmospheric ± 10% 0 to 40mBar max

 

<0.02% signal/mBar

 

0 to 99% RH non-condensing


N.B.  All performance data is based on conditions at 20°C, 50%RH, and 1013mBar


continuous     Long Term Output Drift

Recommended Load

Resistor

Doc. Ref.: 7oxv Issue 1.2

15 to 99% RH non-condensing

<5% signal loss/year 100W

 

 

Performance Characteristics Physical Characteristics


Nominal Range Max Overload

Expected Operating Life

Output Signal

0-25% Oxygen

30% Oxygen Two years in air

0.195 - 0.25mA in air

Storage Life Recommended Storage

Temperature

Warranty Period

Six months in CTL container 0-20°C

 

24 months from date of despatch

(This amounts to a variation of condition


T95

Response Time

£15 seconds

6 of our standard terms and conditions which otherwise apply)

Offset (3mins N2) Temperature Range Temperature Coefficient Absolute Pressure Range

Differential Pressure

Range Pressure Coefficient

Operating Humidity

intermittent

<0.5% O2

-20°C to +50°C

0.2% signal/°C

Atmospheric ± 10% 0 to 40mBar max

 

<0.02% signal/mBar

 

0 to 99% RH non-condensing


N.B.  All performance data is based on conditions at 20°C, 50%RH, and 1013mBar

continuous     Long Term Output Drift

Recommended Load

Resistor

Doc. Ref.: 7oxv Issue 1.2

15 to 99% RH non-condensing

<5% signal loss/year 100W

 

Page 1 of 2 19th October 2004  

 

 

Temperature Behaviour

 

1) Gradual changes

The output of a 7OX-V CiTiceL varies slightly with gradual temperature changes. The behaviour of a batch of 7OX-V sensors is shown opposite. Output was measured at a range of temperatures and expressed as a percentage of the signal at 20°C. The graph shows the mean signal and three times standard deviation.

 

2) Sharp fluctuations

A transient response will occur with sharp fluctuations in temperature. For rapid increases in temperature there is a sharp drop in sensor output, and a sharp increase in output for rapid decreases. These responses are transient and should die away in about 20 seconds. 

 

 

Linearity

The output signal of an Oxygen CiTiceL follows the relationship: S = K loge 1/(1-C)

where:

S = Output signal;

C= Fractional oxygen concentration; K = a constant for the sensor.